===== Rigaku Smartlab II ===== * [[Ligaku SmartLab 1D Reciprocal Space Mapping Error|1D Reciprocal Space Mapping Error]] * [[Optics Allignment(PB) Peak Search Error]] * [[Unknown Alarm|Unknown Error Alarm]] ===== XRD issue History ===== - 장비 도입후 발생한 에러와 문제점들을 모두 기록해두자 (사진과 캡춰화면 원인과 문제점등). * 2022. 06. 21 XRD Intensity가 이전과 다르게 현저히 떨어짐 (Mirror alignment 이후 바로 정상화) * 2022. 06. 13 XRD did not turn on (Be window and related issues), (Be Window 교체 및 청소 06. 21) * {{:be_window_1.jpg?400|}} * 2022. 05. 02 Polishing due to the considerable W peak in measurement (Cu target polishing 05.02) * {{:xrd_drum_head.jpg?400|}} * 2022. 03. 03 Polishing due to the considerable W peak in measurement (Target Drum polishing 03.07) * {{:bro_on_sto_2d_rsm.png?400|}} * 2022. 03. 28 Unknown Error Alarm (필라멘트 교체, Cathod Cleaning 03.30) * {{:xrd_drum_내부.jpg?400|}} * {{:wiki:xrd_fillament.jpg?400|}} * 2022. 04. 12 Optics Allignment error (Is, Zs Cleaning) * 2022. 02. 19 * 2021. 09. 16 * 2021. 08. 18 * 2021. 05. 28 * 2021. 03. 09 * 2021. 02. 17 * 2023. 06. 16 XRD 측정 중 IC server와 교신이 끊김 (IC server 점검, 동일한 증상 발생 시 컴퓨터 바탕화면의 Rigaku service Tools로 log 파일 저장 후 report) * 일정시간 사용 안 할 시 Hard disk 자동 꺼짐을 사용 안 함으로 변경 * Hypix 컴퓨터 내의 Rigaku-Hypix-ini 폴더 교체 (커넥트 실패 에러의 원인으로 예상) * Slit cleaning 진행