사용자 도구

사이트 도구


mpms_squid_magnetometer

차이

문서의 선택한 두 판 사이의 차이를 보여줍니다.

차이 보기로 링크

양쪽 이전 판이전 판
다음 판
이전 판
mpms_squid_magnetometer [2022/10/01 23:42] jyhmpms_squid_magnetometer [2023/06/16 16:10] (현재) yjchoi
줄 6: 줄 6:
 - MPMS 관련 문제, 해결책, 원인에 대해 기록하자. - MPMS 관련 문제, 해결책, 원인에 대해 기록하자.
   * Dewar pressure overshooting and/or undershooting   * Dewar pressure overshooting and/or undershooting
 +  * MPMS Pump 소음과 더불어 Fan의 고장 (2023.06.16)
  
 === History of He regeneration by Evercool system === === History of He regeneration by Evercool system ===
 - He regeneration 이력을 기록하자. - He regeneration 이력을 기록하자.
   * 2022.06.06, target He level: 63%   * 2022.06.06, target He level: 63%
 +
 +=== *** Important things for MPPS prior to do characterizations ===
 +  * There are two fundamentally different control mechanisms used in the system. The normal mechanism is used for all temperatures above about 4.4 Kelvin, while below this temperature, control is achieved by controlling the pressure on a small reservoir of liquid helium. The MPMS control system automatically selects the correct mechanism for the specified temperature.
 +  * Note: The crossover temperature, which can be set from the MPMS control system, depends on the boiling point of helium at ambient atmospheric pressure. Hence, its value will vary with the altitude of the user's site. The correct value for the site is normally set by the Quantum Design service representative when the system is installed and may vary from the sea level value of 4.4 Kelvin.
 +  * The helium consumption in the system is approximately 4.5 to 5 liters/day when running the system below about 200 K, increasing to about 6 to 7 liters/day when running at room temperature. While the MPMS dewar has a 56-liter capacity, the superconducting magnet in the system requires the liquid helium level to be at least 35 to 40 percent to operate. Hence, to operate at the highest fields, the liquid helium should be replenished every 5 to 6 days.
 +  * Verify that a helium-gas supply cylinder is attached to the EverCool manifold supply port with a stainless steel hose and that the cylinder pressure is regulated from 5-30 psi (35-210 kPa).
 +  * Check the Helium Level in the chamber at least over 57%.
 +
  
 === Maintenance checkpoints === === Maintenance checkpoints ===
mpms_squid_magnetometer.1664635348.txt.gz · 마지막으로 수정됨: 2022/10/01 23:42 저자 jyh