mpms_squid_magnetometer [2022/10/08 20:39] jyhmpms_squid_magnetometer [2023/06/16 16:10] (현재) yjchoi
줄 6: 줄 6:
 - MPMS 관련 문제, 해결책, 원인에 대해 기록하자. - MPMS 관련 문제, 해결책, 원인에 대해 기록하자.
   * Dewar pressure overshooting and/or undershooting   * Dewar pressure overshooting and/or undershooting
 +  * MPMS Pump 소음과 더불어 Fan의 고장 (2023.06.16)
  
 === History of He regeneration by Evercool system === === History of He regeneration by Evercool system ===
줄 16: 줄 17:
   * The helium consumption in the system is approximately 4.5 to 5 liters/day when running the system below about 200 K, increasing to about 6 to 7 liters/day when running at room temperature. While the MPMS dewar has a 56-liter capacity, the superconducting magnet in the system requires the liquid helium level to be at least 35 to 40 percent to operate. Hence, to operate at the highest fields, the liquid helium should be replenished every 5 to 6 days.   * The helium consumption in the system is approximately 4.5 to 5 liters/day when running the system below about 200 K, increasing to about 6 to 7 liters/day when running at room temperature. While the MPMS dewar has a 56-liter capacity, the superconducting magnet in the system requires the liquid helium level to be at least 35 to 40 percent to operate. Hence, to operate at the highest fields, the liquid helium should be replenished every 5 to 6 days.
   * Verify that a helium-gas supply cylinder is attached to the EverCool manifold supply port with a stainless steel hose and that the cylinder pressure is regulated from 5-30 psi (35-210 kPa).   * Verify that a helium-gas supply cylinder is attached to the EverCool manifold supply port with a stainless steel hose and that the cylinder pressure is regulated from 5-30 psi (35-210 kPa).
 +  * Check the Helium Level in the chamber at least over 57%.
  
  